Αποτελέσματα Αναζήτησης
The LeadCare II Blood Lead Analyzer displays messages that guide you through the test. Do NOT go to the next step until the message tells you to proceed. The test takes three (3) minutes. When the test is complete, the blood lead level appears on the display.
Intended Use The LeadCare II Blood Lead Analyzer and Test Kit provide a measurement of the amount of lead in a fresh whole blood sample. The LeadCare II Blood Lead Analyzer is intended for in vitro (external) use only.
LeadCare II Blood Lead controls are intended to monitor the accuracy and precision of your blood lead testing using the LeadCare II Blood Lead Testing System. How Often to Test with Controls Test two levels of quality control:1 • on each new lot, or on each new shipment of materials • with each new operator
With the LeadCare II (Waived) Blood Lead Testing System, it is possible to make a rapid diagnosis and to initiate early intervention, since the quantitative analysis of lead in blood is conducted on-site.
The LeadCare II System relies on electrochemistry and a unique sensor to detect lead in whole blood. Most lead is carried in red blood cells. When a sample of whole blood is mixed with treatment reagent, the red blood cells are lysed and the lead is made available for detection.
Lead testing detects the level of lead in the blood (called the Blood Lead Level or BLL). The LeadCare II Blood Lead Analyzer tests the blood lead level and reports it on the screen in three minutes. Results are shown in micrograms (μg) of lead (Pb) per deciliter (dL) of blood.
ISSUE DATE: PRINCIPLE: June 2007. REVISION DATE: May 2012 REVIEWED DATE: May 2018. The LEADCARE II System relies on electrochemistry and a unique sensor to detect lead in whole blood. Most lead is carried in red blood cells.